| Range |
| 20 ohm |
2.0 Kohm |
20 Kohm |
200 Kohm |
2.0 Mohm |
| 1 milliohm |
100 milliohm |
1.0 ohm |
10 ohm |
100 ohm |
|
| Resolution |
| Range |
| 20 ohm |
2.0 Kohm |
20 Kohm |
200 Kohm |
2.0 Mohm |
| 8 mA |
NA |
NA |
NA |
NA |
| 5 mA |
50 µA |
NA |
NA |
NA |
|
| Failsafe Test Current |
| Nominal Current |
The AMPTEC 620VL-T Explosive Safety Igniter Tester - High and Low Ohm Ranges
The AMPTEC 620VL-T Wide Span Explosive
Safety Igniter Tester is a portable digital failsafe ohmmeter which has
been specifically designed for ultra-safe resistance testing on
explosive devices (i.e. squibs, igniters, flares, detonators,
warheads). The 620VL-T is primarily for indoor lab use, the other
AMPTEC 630ES and 640SL (Low Resistance Ruggedized Explosive Safety
meter) are ruggedized water proof versions with similar test
capabilities. The AMPTEC 620VL-T features five way gold panel jacks
that directly accept wiring, spade lugs and banana jacks (i.e. with 4
wire Kelvin Clips). For
a data sheet of all the accessories, test leads, probes and
options that are compatible with the AMPTEC 601ES DMM, 620A-4,
620BH, 620ES, 620GD, 620RK etc select here. The
AMPTEC 620BH uses the same USAF Munition Safety Board Approved Fail
Safe current source for its resistance circuitry. The unit's circuitry
reviewed and approvals are pending by many U.S. Gov. Safety Boards as
well as most branches of the armed forces for specific military
programs.
Wide Span High and Low Ohm Range - Accuracy and Resolution
A common question is “What does
AMPTEC offer to improve the measurement accuracy and resolution with a
low resistance value squib/detonator measurement/application and also
allow isolation testing to 2 Megohms?" This lead to the
development of the AMPTEC 620VL-T Explosive Safety Igniter Tester.
The AMPTEC 620VL-T offers the following fullscale ranges - 20 Ohms, 2.0
KOhms, 20 Kohms, 200 Kohms and 2.0 Megohms. The 20 Ohm fullscale range provides a four-wire Kelvin resistance measurement
from 0.000 ohms to 19.999 ohms. The 20 Ohm range also offers a low 1 milliohm resolution. AMPTEC RESEARCH
offers an optional Four Wire Kelvin (OP300) Lead set that eliminates
the effects of test lead length resistance and contact resistance
errors.
Isolation Testing to 2.0 Megohms
The 2.0 Megohm range on the AMPTEC 620VL-T provides a high resistance range for determining open circuit or "electrical isolation between any two contact test points or wires" (>2.0 Megohms) status on ordnance wiring and detonators. See the AMPTEC 620BH if you need to measure isolation resistance to a higher level up to 20.0 Megohms (620BH).
Incredibly Safer than a Commercial DMM
Commercial Multifunction DMM’s have
many modes (i.e DC Voltage, AC Voltage, DC current, AC current,
frequency, shift keys, Capacitance and more..) that can accidentally
mis-measure igniter circuit resistance. If the DMM has an
electrical fault the igniter circuit may accidentally fire. The
620VL-T Igniter Tester only measures resistance, it is not a complex
multifunction DMM. DMMs also aren't optimized for squib resistance
measurements, most squibs are around 1.0 Ohm and the AMPTEC 620VL-T
Explosive Safety Igniter Tester offers a 20 Ohm full scale range.
Polyurethane Epoxy Conformal Coated Circuit Board and More ...
A common question we get from safety
engineers is "What sort of electrical and mechanical
measures are taken with the AMPTEC 620BH to prevent an internal
loose wire, metal screw, or washer inside from floating around
and shorting the battery output to the terminal output posts?".
The entire PCB (except for the calibration trim pots) is
non-conductive polyurethane epoxy coated, to make it extremely
difficult to get an accidental conductive pathway. For starters,
the 620VL-T Explosive Safety Igniter Tester fundamentally uses the USAF
Non-Nuclear Safety Board approved AMPTEC 620A-4 Igniter Tester
Failsafe ciircuitry. The 620BH has the added circuitry of the 2.0
Ohm range which has also been approved by the US Dept. of Energy (High
Explosive Safety Committee) in the form of the AMPTEC 630ES and AMPTEC
640SL. The ultimate electrical safety analysis for failsafe
current relies on Ohm's Law for determining Imax based upon a Vmax (~
4.8 VDC battery value) divided by the equivalent circuit current
limiting resistance (R-limit) for the particular version being
considered. The AMPTEC 620A circuit schematics are available upon
request for reviewing a version's maximum failsafe current level.
Kelvin Clips , Probes and Test Lead Accessories
For
a data sheet of all the accessories, test leads, probes and
options that are compatible with the AMPTEC 601ES DMM, 620A-4,
620BH, 620ES, 620GD, 620RK etc select here.
|
Option "403" 4 Terminal Kelvin Mini-Probes
- (photo shown above) Each probe has two spring loaded, gold plated,
steel with 0.18" separation. (one red and one black handheld probe).
Excellent general purpose Mini-Probe, and along with the Option "300"
Kelvin Clip Leads easily fit inside the AMPTEC 620VL-T Test Lead Pouch
(OP110).
|
 
Option "403" Kelvin Mini-Probe Set
|
|
Option "300" Kelvin Clip Test Leads - Gold Plated Four Terminal Kelvin Clip Lead Set open to ½" for connection to wires, lugs, etc.
|

Option "300" Kelvin Clip Test Leads
|
|
Option "401" Single Pointed Handheld Probe Set
- Ideal for probing into recessed squib wire sockets, and onto larger
metal surfaces (flat or round) such as flares, rounds and shells, that
do not easily accept test lead clip attachment.
|

Option "401" Test Probe Lead Set
|
 |
Optional 24/7 Continuous Isolated Operating Power - 24 Hours a Day
To continually use the AMPTEC
620VL-T,especially in an automated application with a computer would
eventually drain any battery powered unit. Option "247" allows the
AMPTEC 620VL-T Igniter Tester to be safely operated 24 Hrs/Day 7
Days/Week continuously. It is based upon the same "Hospital
Safety Isolation" technology used in monitoring and diagnosing human
patients while wired/connected to 120 VAC powered electrical instrumentation.
This Ultra High Line Isolation DC/DC convertor based option is rated to
3,000 volts isolation._
|
| Specs: |
| Accuracy: |
(180 days 25 ºC ± 10
ºC) ± 0.02% of reading ± 0.02% of range for the 20
ohm thru 20Kohm ranges, ± 0.02% of reading and ± 0.05% of
range for 200Kohm range, ±1.0% of reading ± 0.2% of
range for the 2 Megohm range.
|
| Failsafe Current: |
< 8mA - 20 ohm range, |
| Display: |
4½ digits (19999) Superbright LED - 20,000 counts. Display is visible in direct sunlight. |
| Terminal Configuration: |
Four-wire Kelvin |
| Resistance Ranges |
20 Ohm, 2.0 KOhm, 20 Kohm, 200 Kohm, & 2.0 MegOhm range |
| Overload Indication: |
Display Flashes |
| Ohms Measurement: |
4 Wire Kelvin with 5Way Binding Posts on the front panel
|
| Operating Temperature Range |
0 ºC to 50 ºC
|
| Display Update Rate |
Approximately 3 per second |
| Power: |
Rechargable Ni-Cad batteries (5.0 Ahr) & Charger (115VAC 50-60Hz) |
| Size: |
235mm/9" D x 254mm/10" W x 76mm/3" H
|
| Weight |
1.6 KG/3.5lbs NET 3 KG/6.5 lbs shipping
|
|